Gas Chromatograph
Hargamesin, jualmesin, distributormesin, carimesin, AlatUji Laboratorium, mesinplastik, pengolahankaret, DaurUlang, Pertanian, Printing, Tekstil, karetLatex
Hargamesin, jualmesin, distributormesin, carimesin, AlatUji Laboratorium, mesinplastik, pengolahankaret, DaurUlang, Pertanian, Printing, Tekstil, karetLatex
WhatsApp
085222110091
WhatsApp
085277110091
  • Distributor Mesin Industri dan Alat Uji Laboratorium

ONE-STOP SOLUTION FOR YOUR INDUSTRY and LABORATORIES

Hargamesin, jualmesin, distributormesin, carimesin, AlatUji Laboratorium, mesinplastik, pengolahankaret, DaurUlang, Pertanian, Printing, Tekstil, karetLatex Gas Chromatograph Gas Chromatograph

Gas Chromatograph

GC122

Application Areas

§  Trace toxics analysis, monitoring and research in environmental protection, atmosphere and water source pollution.

§  Biochemistry, clinic application, pathology and toxicology research.

§  Microorganism analysis and research in food ferment, and microelements analysis in beverage.

§  Intermediate and finished product analysis in Chinese and western medicine materials.

§  Crude oil composition analysis and control in petroleum processing, petrochemical industry, petroleum geological industry, and geological prospecting research;

§  Component research and production control in organic chemistry and organic synthesis.

§  Analysis and research in sanitary quarantine, labor protection, and hazard detection.

§  Monitoring, control and research in cuttingedge science and military technologies.

 

 

Features

·Newly designed, high performance HRGC

·Double FID detectors in the basic type with wide accessory options such as GDl22TCD thermal conductivity detector, GDl22ECD electron capture detector, GDl22NPD nitrogen and phosphorous detector, GDl22FPD flame photometric detector, converter, gas sample inlet valve, etc.

·In the basic type, the instrument has dual packed columns, double samplers and dual gas path system structure, which can effectively restrain baseline fluctuation and drift in real time, suitable for temperature programming analysis. The instrument can work with three types of detectors at the same time. With different detector combination in different application, the instrument can support detectors working in either cascade or paralleled connection.

·The instrument in the basic type is equipped with completed capillary analysis system and independent capillary split sampler, which includes diaphragm cleaning, split flow and auxiliary gas adjustment. It supports capillary split / split less sampler, cold on column sampler, and 0.53mm large diameter capillary column direct sampler. It’s suitable for all kinds of capillary column of various diagrams, speculations, and lengths, and made of glass or flexible silicon glass.

·The gas path of the instrument adopts high precision scale flow control valve, unloading valve and needle valve. Its easy to adjust flow capability and has better reproducibility and stability.

·The gas path of the instrument adopts high precision scale flow control valve, unloading valve and needle valve. Its easy to adjust flow capability and has better reproducibility and stability.

·Packed column: on column sample injection, instantaneous vaporizing sample injection, gas sample injection

·Capillary column: split injection, split less injection, large diameter column direct injection

· Newly developed computer aided temperature control system. It can manage the temperature control of six path heating areas at the same time, including column, ion chamber, thermal conductivity detector, capillary column sampler, packed column sampler auxiliary and standby, with high precision (less than ±0.1oC) and wide range (up to 400oC), and temperature overload and deviation is less than 2oC or even lower.

·The column has five phase programmed temperature control, and the highest heating rate is up to 40oC/min.

·The system has such functions as switch off protection, temperature limit setup, temperature scanning, rapid automatic cooling, i.e. back opening door, and etc.

·Open-end computer system can apply RS232 interface and N2000 chromatogram workstation to achieve single path communication control and data processing.

·Large volume full lens stainless column (350mm×280mm×300mm) can support.

Pesan Sekarang
Order lewat Whatsapp
CONTACT US

SALES OFFICE ADDRESS
Kawasan Perkantoran Metro Indah Mall blok D-35
Jalan Soekarno Hatta Bypass Area
BANDUNG City - JAWA BARAT Province – INDONESIA
 
PHONE
OFFICE : 022.7536349 (HOTLINE)
SALES : 0852.22.110091 / 0852.77.110091
E-PROC / PENGADAAN : 0852.11222.091
TECHNICAL / OPERATIONAL : 0852.118000.91

 

PROCUREMENT (PENGADAAN) :
eproc@valtekindo.co.id
 

CUSTOMER SERVICE & AFTER SALE SERVICE :
cs@valtekindo.co.id
 

SALES & TECHNICAL DEPARTMENT :
sales@valtekindo.co.id
 


 

©2020 PT.VALTEKINDO GLOBAL INTERTEK All right reserved   Developed by javwebnet